27
10 CONFIGURATION
The operative configuration of the unit can be changed using the below indicated procedure.
Switch the instrument -
Off
-
Keep the <
ENTER
> key pressed and press the <
Trigger
> push button until the instrument displays the
indication relevant with the technical unit selection.
10.1 Technical Unit selection
the display will indicate one of the following technical units (Celsius, Fahrenheit and Kelvin):
e
Set
Lock
e
Set
Lock
e
Set
Lock
Press the <> or <> key if you require to select a new technical unit
Press the <
SEL
> key to select the new configuration step or press the <
Trigger
> to switch the instrument -Off-. The
new selection will be automatically memory stored.
10.2 Thermocouple input enabled
Press the <
SEL
> key until the following indication is displayed :
e
Set
°C
Lock
or
e
Set
°C
Lock
Press the <> or <> key to enable (or to disable) the thermocouple auxiliary input operation.
Press the <
SEL
> key to select the new configuration step or press the <
Trigger>
to switch the instrument -Off-. The
new selection will be automatically memory stored.
10.3 Temperature Scale selection
Press the <
SEL
> key until one of the following indication is displayed :
e
Set
°C
Lock
or
e
Set
°C
Lock
Press the <> or <> key to select the required International Temperature Scale.