Omega PCL340 Manual de usuario

Busca en linea o descarga Manual de usuario para Control remoto Omega PCL340. Omega PCL340 User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir

Indice de contenidos

Pagina 1 - User’s Guide

omega.com e-mail: [email protected] latest product manuals:omegamanual.infoUser’s GuideM-4292/1109PCL340 SeriesShop online at

Pagina 2

6Figure 3 DisplayTable 3 Display FunctionsNo. Name Description 1 Primary Parameters Indicates what is being measured. 2 Span Indicator Indicate

Pagina 3

7OTHER DISPLAYSWith a single display the following modes are available:P[1] = Pressure on left side sensor.[EXT] = Pressure with external pressure mod

Pagina 4

8Figure 4 Menu Map

Pagina 5 - 1. Introduction

92.2 Using the BacklightThe backlight is controlled by the dedicated backlight key. It toggleson and off when the key is pressed; this is one of the

Pagina 6 - Symbol Description

102.4.1 Setting the ContrastFrom the Contrast Main Menu choose the CONTRAST option toaccess the Contrast Adjustment Menu. Use the arrow keys to adjust

Pagina 7

112.4.4 Setting AutoShut-off ParametersThe calibrator can be set to automatically shut-off after a selectednumber of minutes; this function can also b

Pagina 8 - 2. Calibrator Interface

12configurations of the other currently active displays, if theconfigurations are in conflict the recalled display’s configuration ismodified to avoid

Pagina 9 - 2.1 Calibrator Display

13including gauge, vacuum, differential and absolute. The moduleswork seamlessly with the calibrator. Simply plug them into theinterface and select

Pagina 10

Figure 75. Measuring VoltageTo measure voltage use the input terminals in the front of thecalibrator. Select the Volts function on one of the display

Pagina 11

6. Performing a Pressure Switch TestFigure 10To perform a switch test, follow these steps:1. Change the setup to Setup 4 (default switch test).Setup

Pagina 12 - Menu Map

Servicing North America:USA: One Omega Drive, Box 4047ISO 9001 Certified Stamford CT 06907-0047Tel: (203) 359-1660 FAX: (203) 359-7700e-mail: info@ome

Pagina 13 - 2.3 Using the Zero Function

5. Apply pressure with the pump slowly until the switch opens. Important NOTE: In the switch test mode the display update rate isincreased to help ca

Pagina 14

179. Press the “NEW TEST” option to clear the data and performanother test.10. Press the “DONE” option to end the test and return to thestandard press

Pagina 15

187.2 Calibrating a Pressure-to-Current TransmitterTo calibrate a pressure-to-current transmitter (P/I), perform thefollowing steps:1. Connect the cal

Pagina 16 - 3. Measuring Pressure

19Figure 12.Example:Suppose a pressure transmitter under test is 30 psi (2 Bar) Full Scaleand outputs a corresponding 4 to 20 mA signal. The user can

Pagina 17 - 4. Measuring Current

204. LOOP POWER can be toggled on/off, select NEXT when done.5. Use SELECT to toggle through the UNIT options, and select NEXTto move on.6. Use the ↑

Pagina 18 - 5. Measuring Voltage

21Note: The 0% and 100% point will be saved in non-volatile memoryuntil they are changed again by the user for the internal sensors, andexternal press

Pagina 19

22Setup 2: The upper display is set to [EXT] mode and the middle isset to mA, lower is off. Setup 3: The upper display is set to [P1] mode and th

Pagina 20

2310. Remote Operation10.1 Remote InterfaceThe calibrator can be remotely controlled using a PC terminal, or bya computer program running the calibra

Pagina 21 - 7. Calibrating Transmitters

24To set up remote operation of the calibrator on the Windows HyperTerminal, connected to a COM port on the PC as in Figure 23, usethe following proce

Pagina 22 - 7.3 Percent Error Function

25The calibrator may be controlled using commands and queries. Allcommands may be entered using upper or lower case. Thecommands are divided into the

Pagina 23

1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11.1 Contacting OMEGA . . . . . . . . . . .

Pagina 24

IntegerFor most computers and controllers they are decimal numbersranging from -32768 to 32768. For example:FAULT? could return 110Refer to the Error

Pagina 25 - 9. Factory Setups

10.5 Remote Commands and Error CodesThe following tables list all commands, and their descriptions, thatare accepted by the calibrator. Table 5: Commo

Pagina 26

28ST_START Starts a switch test ST_OPEN? Returns pressure value at which the switch opened ST_CLOSE? Returns pressure value at which the switch

Pagina 27 - 10. Remote Operation

29Table 8: Error CodesError Number Error Description100 A non-numeric entry was received where it should be anumeric entry 101 Too many digits ent

Pagina 28 - 10.4 Using Commands

3010.6.2 Calibrator CommandsDAMPTurns the dampening function on or off.For example:If you send DAMP ON this will turn the dampening functionon. DAMP?R

Pagina 29

31ERROR _LOOPTurns loop power on or off in percent error mode.For example:To set loop power on send ERROR_LOOP ON.ERROR _LOOP?Returns the current stat

Pagina 30

FUNCSets the display indicated in argument one to the functionindicated in argument 2.For example:To set the lower display to P1 Pressure mode send FU

Pagina 31

LO_ERRSets the 0% point for the percent error mode calculation in thecurrent engineering units.For example:To set the 0% point to 20 psi send LO_ERR 2

Pagina 32 - Table 7: Parameter units

ST_DEAD?Returns deadband of the switch in the current pressure units.VAL?Returns the value of any measurement taking place on the upperand lower displ

Pagina 33 - 10.6 Entering Commands

11. Specifications (18 °C to 28 °C unless otherwise noted.)GeneralInstrument Setup Recall 5; last used on power-upEnvironmentalOperating Temperature

Pagina 35

12. Maintenance12.1 Replacing BatteriesReplace batteries as soon as the battery indicator turns on to avoidfalse measurements. If the batteries discha

Pagina 38

WARRANTY/DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials andworkmanship for a period of 13 months from date of

Pagina 39 - Accuracy

Where Do I Find Everything I Need for Process Measurement and Control? OMEGA…Of Course!Shop online at www.omega.comTEMPERATUREMUThermocouple, RTD &

Pagina 40 - 12. Maintenance

1. IntroductionThe OMEGA PCL340 Series is designed to be a simple to use yetvery versatile pressure calibrator. Its internal pressure sensorcombined w

Pagina 41

Symbol DescriptionDouble InsulatedElectric ShockFusePE GroundHot Surface (Burn Hazard)Read the User’s Manual (Important Information)OffOnThe following

Pagina 42

3• Select the proper function and range for your measurement.• Make sure the battery cover is closed and latched before youoperate the calibrator.• Re

Pagina 43 - RETURN REQUESTS/INQUIRIES

2. Calibrator Interface Figure 1 shows the location of the process measurement inputs,while table 1 describes their use.Figure 1 Process Measurement I

Pagina 44 - OMEGA…Of Course!

Figure 2 KeypadTable 2 Key FunctionsNo. Name Description 1 Function Keys These keys are used in various ways, primarily toconfigure the calibrator