CAUTION!
Remove power to unit before wiring
input and output connections.
Follow instructions carefully to avoid
personal injury.
Omega CDTX-90 Series Conductivity/Resistivity Transmitter
Contents
1. Installation
2. Speci cations
3. Electrical Connections
4. Menu Functions
2. Speci cations
General
Compatible electrodes: Omega CDCE-90 Series Standard and
Certi ed Series Conductivity/Resistivity Electrodes
Accuracy: ±2% of reading
Enclosure:
Rating: NEMA 4X/IP65 front
• Case: PBT
Panel case gasket: Neoprene
Window: Polyurethane coated polycarbonate
Keypad: Sealed 4-key silicone rubber
Weight: Approx. 325g (12 oz.)
Display:
Alphanumeric 2 x 16 LCD
Contrast: User selected, 5 levels
Update rate: 1.8 seconds
Electrical
Power: 12 to 24 VDC ±10%, regulated, 80 mA max.
Sensor input range:
Conductivity: 0.055 to 400,000 μS
• Resistivity: 10 KΩ to 18.2 MΩ
( Measurements from 10 MΩ to 18 MΩ (0.055 μS to 0.1 μS) must
be performed in solution temperatures from 20 °C to 100 °C. )
TDS: 0.023 to 200,000 ppm
Temperature: PT 1000, -25 to 120°C (-13 to 248°F)
Current output:
4 to 20 mA, isolated, fully adjustable and reversible
Max loop impedance: 50 Ω max. @ 12 V
325 Ω max. @ 18 V
600 Ω max. @ 24 V
Update rate: 200 mS
Accuracy: ±0.03 mA @ 25°C, 24 V
Relay outputs
(CDTX-90-2 includes 2 mechanical SPDT contacts):
Maximum voltage rating:
5 A @ 30 VDC, or 5 A @ 250 VAC, resistive load
• Programmable for:
High or Low setpoint with adjustable hysteresis
Pulse operation (max. rate 400 pulses/min)
Open-collector output, optically isolated:
50 mA max. sink, 30 VDC maximum pull-up voltage.
• Programmable for:
High or Low setpoint with adjustable hysteresis
Pulse operation (max. rate 400 pulses/min)
Environmental
Operating temperature: -10 to 70°C (14 to 158°F)
Storage temperature: -15 to 80°C (5 to 176°F)
Relative humidity: 0 to 95%, non-condensing
Maximum altitude: 2000 m (6562 ft)
• Insulation category: II
• Pollution degree: 2
Standards and Approvals
• CE
• Immunity: EN50082-2
• Emissions: EN55011
• Safety: EN61010
Manufactured under ISO 9001 and ISO 14001
U.S.A. Patent # 5,708,363
1. Installation
CDTX-90 Series transmitters are available in two styles: panel mount and eld mount. The panel mount is supplied with the
necessary hardware to install the transmitter. This manual includes complete panel mounting instructions.
Field mounting requires a separate mounting kit. The FP90UM Universal kit enables the transmitter to be installed virtually
anywhere. Detailed instructions for eld installation options are included with the FP90UM Universal kit.
1.1 Panel Installation
1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel cutout, an adhesive
template is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch.
2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE.
Press quick-clips outward and remove.
Conductivity/Resistivity
62.50 uS/cm
25.0° C
ENTER
92 mm
(3.6 in.)
97 mm
(3.8 in.)
56 mm
(2.2 in.)
41 mm
(1.6 in.
)
Optional
Rear
Cover
FRONT VIEW
96 mm
(3.8 in.)
96 mm
(3.8 in.)
Field &
Panel Mount
SIDE VIEW
Field Mount with
Universal kit
106 mm
(4.18 in.)
82 mm
(3.23 in.)
Panel Mount with
NEMA 4X rear cover
SIDE VIEW
Field Mount with
Integral kit
96 mm
(3.8 in.)
107 mm (4.2 in.)
42 mm
(1.7 in.)
65 mm
(2.5 in.)