-127-
22. SPECIFICATIONS
lMessage : Printing of message in message No. selected
Numbers of message: Max. 5
Message programming: Max. 15 digits
Chart paper: Depending on models
Model System
Printing
width
To ta l
width
To ta l
lengt
h
RD200 Folding type 100mm 114mm 10m
RD2800 Folding type 180mm 200mm 20m
Chart speed: 1 to 600mm/h, 1 to 200mm/h
Default … RD200: 20mm/h
RD2800: 25mm/h
Chart speed accuracy: Within ±0.1%, For the chart scale
after feeding 1000mm or more.
Skip function: No display or printing is executed for the
channels for which no range is programmed.
Time axis synchronization:
On or off selectable with key operation
Subtract printing:
Printing of difference between measured value and
reference channel (or reference value)
Pen lift: Pens are automatically lifted up when printing is off.
3 Indication and Display
Trace printing indication: Bargraph display for each channel
Model
No. or
segment
Channel color segment
RD200 51 5-segment each
RD2800 101 10-segment each
Digital display: 16- and 7-segment LCD with cursor
16-segment … Character height: 7.5 mm, orange
7-segment …. Character height: 6.5 mm, white
Model 16-segment 7-segment
RD200 1 digit 20 digits
RD2800 1 digit 30 digits
Display items: cSimultaneous display of measured values
for all channels d Year/month/day (only RD200) e
Time f Chart speed g Alarm status
Display switching: Each time is pressed, the
display item changes.
RD200 .… c def g
RD2800 … cef g
Measured value display: -9999 to 99999 for each channel
(Decimal point can be placed at desired position with
the scale programming )
Status: Following 6 status lamps are available.
Display text Condition to light Color
RECORD ON Printing ON Green
KEY LOCK Key lock Blue
ALARM Alarm occurrence Red
CHART END Just before chart paper ends Red
FAIL
Hardware related to servo
mechanism abnormal
Red
POC Time axis sync. ON Blue
Chart illumination: CFL
4 Alarm
Alarm point: Numbers of channel x numbers of level (4)
Alarm display: Alarm status lamp lights and the
measured value blinks at alarm activated channel.
Alarm display screen indicates alarm channel and
alarm type at level digit.
Alarm type: Selectable from the following 10 types for each
alarm point
H: High limit L: Low limit
Absolute
value
E: High limit with standby
F: Low limit with
standby
Rate-of-
change *1
U: Increase limit D: Decrease limit
B: Differential high limit S: Differential low limit
Difference
*2
J: Differential high limit
with standby
K: Differential low limit
with standby
*1: Variation range per unit time (Note)
(Note) Measuring cycle x measuring count (1 to 20)
*2: Difference between two channels
Alarm deadband: 0.1 to 9.9% of scale programming range
Alarm output: Option
5 Programming and Operation
Key types: See Section 7.1.
Key functions: See Section 7.2 .
Basic operation: Basic operation with keys
c Printing ON/OFF d Chart paper feeding
e Operation screen switching
Operation: Operation with keys
Selection of c Digital data printing *1 d List
printing *1 e Message printing *1 and f Printing
format*2
*1: Operation with remote contacts (option) is also
available
*2: Only available with option
Pen change mode: Pen moves to the position to beeasily
changed.
Standard programming:
Range/printing range, °C/°F computation, chart speed,
time, scale, skip, subtract printing, alarm, alarm
deadband, periodic time printing, engineering unit, tag,
message, burnout, passcode/key lock, input filter
Optional programming:
Alarm output…Relay No., AND/OR, output mode
Remote contacts…Terminal allocation for operation, 3
chart speeds, operation record position
Printing format……Automatic range-shift printing,
compressed/expanded, zone printing
Communications interface…Communications protocol,
communications specifications
Math expression…Selection of types, parameters
Totalization…Starting time, interval
Engineering port: All parameters can be programmed with
engineering software “PASS” (option) installed on a
personal computer