-41-
8. BASIC PROGRAMMING
8.2 Range/Printing Range
3 Programming Flow Chart
<Example> Range No. for channel 3: 22, RJ: Enable, Printing range: -50 to 450°C
Operation screen
[Check]
+
[To be ready for programming]
[Selecting channel]
→
[Selecting range No.]
→
[Programming RJ selection]
[Programming printing range]
Error message appears?
Temporarily stored
Programming other channels?
+
Stored
Channel 1 range.
Other channels can be checked with
or .
• Cursor and programming lamp light.
• Cursor moves rightward when
pressing a numeric key.
Cursor moves rightward.
Cursor moves leftward.
Press or , and select
[3], then press .
Press and then press
again (range No. 22). See the next
page for the range No. table.
RJ (Reference Junction compensation):
0 → Disable*
1 → Enable
*For all inputs other than
thermocouple input, program to “0”.
Adding/deleting decimal point: Move
the cursor rightward and press
or
.
• Programming “to”: Press and
simultaneously.
• Decimal point is not included in the
number of programming digits.
An error occurs if a value exceeds the
programming range or unnecessary
space is entered.
Display moves to the next channel for
programming.
Store the [temporarily stored]
programmed values into memory.
Memory colon blinks and the
programming change mark is
printed.
YES
NO
YES
NO
Cursor
Channel
Range No.
RJ selection
Printing range
*
Channel
* Unused digits should be filled with spaces.
RANGE
END
TAG
TAG
CHART
Function keys
Keys other than
℃
END
TAG
TAG
COPY
=
・
CLOCK
-
-
Comentarios a estos manuales